エラー ID: 34224
The home position of the metering plunger is not found or the plunger is unable to move.
考えられる原因 | 対策 | |
|---|---|---|
1 | Blockage in the flow path. | Check if there is a significant pressure difference (> 10 – 15 bar) between the mainpass and the bypass position of the injection valve. If yes, systematically replace the components of the mainpass: Needle seat, Needle, Rotor seal, Sample loop |
2 | Potential hardware error. | Agilent サービス担当者までお問い合わせください。 |