エラー ID: 1018
The visible lamp current is missing.
The processor continually monitors the lamp current during operation. If the current falls below the lower current limit, the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | Lamp disconnected. | Ensure the lamp connector is seated firmly. Ensure the lamp is connected. |
2 | Defective visible lamp. | Exchange the visible lamp. |
3 | Defective connector or cable. | Agilent サービス担当者までお問い合わせください。 |
4 | Defective power supply. | Agilent サービス担当者までお問い合わせください。 |
base-id: 3359321355