The SFC Control Module has a waste bottle located outside of the cabinet. It can be located anywhere easily accessible and visible within the range of the supplied tubing. The waste bottle serves multiple purposes and collects liquid waste from multiple sources. The primary purpose is to separate the gaseous and liquid waste from the outlet of the BPR (system) in such a manner that the gaseous waste can be appropriately vented outside of the lab environment. The waste bottle has input and output ports located above any collected liquids. The mixed stream enters the waste bottle and the gaseous stream exits from the spout.
NOTE
Proper system operation requires adequate space in the waste bottle to allow gaseous exit. It is the responsibility of the operator to ensure that the waste bottle is empty before beginning operation of the SFC control module, and to monitor and empty the waste bottle as needed during usage. This is not a warning.
WARNING
Exposure to toxic substances
Locate and assemble the waste bottle and Tygon vent tubing. The vent tubing can be placed over the spout on the top of the waste bottle. Route the Tygon tubing to an appropriate vent. The system must be actively vented.
Locate the SFC control module Waste line. Insert the free end through a hole in the top of the waste bottle cap. Insert the tube half-way into the waste bottle. Connect the fitting end to the outlet union on the BPR. This union uses a 10-32 CPI fitting. Tighten snugly.