| The Sampler control parameters are in the following
sections: Missing Vial Mark the
Ignore missing vial check box to
specify that, if a vial is missing, the injector ignores it and continues with
a 6-second dummy run. The message "Missing vial <x>" is logged, and the
system continues with the next injection.
Illumination Toggles the
illumination of the sample area, On or Off. At Power On The
section is available when a cooler/thermostat is installed and configured. Mark
the Turn on Thermostat check box
to specify that the cooler/thermostat is switched on automatically when the
instrument is switched on.
Thermostat The section is
available when a cooler/thermostat is installed and configured and the Constant
temperature mode is selected. Select On to switch on the cooler/thermostat. Specify the required
temperature in the adjacent field. The specified temperature must be at least
5 °C below ambient for proper
temperature control. Select Off to
switch off the cooler/thermostat.
Automatic Turn On You can set a date and time at which the cooler/thermostat switches on
automatically.
Pump connected to Sampler
Use this section to specify the pump that is used with the
Sampler. If more than one pump is configured, display the drop-down list and
select the appropriate pump from the list. It is critical for Feed
Injection.
Enable Analysis
This feature
requires LC & CE Drivers A.02.19 or newer and is only available for the
Sample Thermostat. With this function, you can specify if the analyses should
start With any temperature or
only when the Temperature is
within ± 2 °C range of the setpoint temperature.
NOTE: The Enable Analysis section is disabled when Not
controlled is selected in the Temperature
section.
|