For best performance of the Sampler
When using the sampler in a system with a vacuum degassing unit, shortly degas your samples before using them in the sampler.
Filter samples before use in an InfinityLab LC Series system. Use 5067-6189 (InfinityLab Inline Filter (0.3 µm)) for inline filtering.
When using buffer solutions, flush the system with water before switching it off.
Check the sampler plungers for scratches, grooves, and dents when changing the piston seal. Damaged plungers cause micro leaks and will decrease the lifetime of the seal.
Solvent Information: Observe recommendations on the use of solvents, see Solvent Information.
Priming and purging the system: When the solvents have been exchanged or the system has been turned off for a certain time (for example, overnight) oxygen will re‐diffuse into the solvent channel. Therefore priming and purging of the system is required before starting an application.