When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 8710-2392 | 4 mm Hex key |
1 | ![]() | 01018-23702 | Insert tool OR |
1 | ![]() | G4226-43800 | Seal insert tool |
Prerequisites
|
In the Local Controller start the maintenance mode and select Change Metering Device function.
OR
In the Agilent Lab Advisor software select Service & Diagnostics in the system screen Maintenance Positions > Change Metering Device , click Start and wait until the metering device is in maintenance position.
Open the front door.
Take the metering device. Push against the rear side of the metering device and rotate a quarter left to release the bayonet lock.
If dirty: Clean the piston with an appropriate solvent. If scratched: Replace the piston by a new one.
Carefully remove the metering seal using the steel side of the insert tool. Clean the chamber with an appropriate solvent and ensure that all particulate matter is removed.
base-id: 5702212107
id: 9007204956953099