To achieve best performance of the module, follow these rules.
When using the module in a system with a vacuum degassing unit, shortly degas your samples before using them in the module.
Filter samples before use in a Agilent InfinityLab LC Series system. Use 5067-6189 (InfinityLab Inline Filter (0.3 µm)) for inline filtering.
When using buffer solutions, flush the system with water before switching it off.
Check the module plungers for scratches, grooves and dents when changing the piston seal. Damaged plungers cause micro leaks and will decrease the lifetime of the seal.
Solvent Information - Observe recommendations on the use of solvents, see Solvent Information.
Priming and Purging the System - When the solvents have been exchanged or the system has been turned off for a certain time (for example, overnight) oxygen will re‐diffuse into the solvent channel. Therefore priming and purging of the system is required before starting an application.