The Assist Hub (G7180A) power supply has wide ranging capability. It accepts any line voltage in the range described in Specifications of the G7180A InfinityLab Assist Hub. Consequently there is no voltage selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.
WARNING
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power line at any time.
WARNING
Incorrect line voltage at the module
Shock hazard or damage of your instrument can result if the devices are connected to line voltage higher than specified.