The ELS Detector can store in memory a series of time-based events, within a single timetable. This timetable allows the operational settings of the ELSD to be changed in real time during a run.
The evaporator temperature (G7102A with cooled option), gas flow, PMT gain (G4260B), and smoothing parameters can all be configured within this timetable to change during a sample injection.
The timetable can be used to program the gas flow, to compensate for the change in ELSD response across a solvent gradient.
The single timetable, stored onboard the ELSD, is only customizable using the ELSD Dimension software, which can be purchased from Agilent (PL0890-0375 (ELSD Dimension Software) ). A trigger cable has to be ordered separately. Use G4260-63002 (Infinity II/III Trigger Cable for Dimension Software) for an Infinity II/III system, and PL0890-0350 (Remote start cable) for a third-party system.