Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Data acquisition leads not connected | Ensure connectors to computer or integrator are securely connected. |
2 | Light source inactive | Check LED or LASER intensity is 100 %. Check LED or LASER is functioning correctly, by stopping solvent flow, cycling the power. Then reading the offset value in RUN mode should be below 130 mV. |
3 | Output below 0 mV | Stop pump flow and A/Z without liquid flowing. |
4 | Instrument in STANDBY mode | Select RUN mode. |
5 | Nebulizer or nebulizer inlet tube blocked | Manually syringe water into ELSD front Inlet port to remove obstruction. |