Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Offset too high or output unstable due to impurity in mobile phase | Stop pump flow and switch off unit. Restart unit and A/Z without liquid flowing. Optical section contaminated and requires cleaning. Please contact your Agilent service representative. |
base-id: 4344409995