Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Partial blockage in nebulizer or nebulizer inlet tube | Pump a 50/50 water/acetone mixture into ELSD at highest flow rate possible (do not exceed 5 mL/min) for 16 h. |
2 | Internal solvent trap is empty | Fill the front solvent trap with liquid until any excess flows out through front drain tube. |
3 | Gas pressure too low | Ensure inlet gas pressure >60 psi. |
4 | Laser power | Verify by switching the unit from STANDBY to RUN and check offset. |
5 | Optical chamber is contaminated | Clean contamination from widows in optical chamber. |
6 | Diffuser saturated with solvent | Stop the eluent flow and increase the evaporator temperature to maximum. Increase the flow rate to 2.8 SLM and wait 1 h. |