Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Inconsistent nebulization | Nitrogen is the recommended gas - others can be used but may not nebulize as efficiently. |
2 | Incorrect gas being used | Change gas to nitrogen or evaluate different nitrogen sources. |
3 | Poor regulation of inlet gases | If using bottled gas, check that gas regulator is functioning correctly and giving consistent flow. Alternatively, use a pulse dampener. |
4 | Insufficient smoothing | The broader the peaks, the higher the smoothing value is required. Increase smoothing to 50 for Flash separations. |
5 | Sample precipitation during nebulization | Reduce sample concentration or inject same loading in larger volume. Check sample solubility in mobile phase eluents. |
6 | Inconsistent pump flow rates | See remedy for pump pulsation in the baseline noise section. |
base-id: 4344291211
id: 9007203599032203