Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Ambient lab temperature too high | Move detector to laboratory where ambient temperature is <25 °C. |
2 | Faulty Peltier cooler unit | Please contact your Agilent service representative. |