Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Inefficient evaporation | Increase the evaporator temperature and/or gas flow. |
2 | High concentration of nonvolatile buffer or stabilizer | Use a lower concentration of stabilizer, unstabilized solvent or, a more volatile buffer (ammonium acetate or ammonium formate). |
3 | Contaminated diffuser | Perform cleaning procedure. |
4 | Optics Heater failed | Please contact your Agilent service representative. |
base-id: 4344129163