Error ID:
Probable cause | Suggested actions | |
---|---|---|
1 | Particulate matter in the gas supply | Filter the incoming gas, or change the supply. |
2 | Column shedding | Replace column or fit an inline filter with a 0.2 µm membrane filter directly after the column. |
3 | Poor nebulization | Check solvent flow rate into ELSD is constant. Check inlet gas flow is >60 psi and stable. |
4 | Insufficient evaporation | Increase the temperature of the evaporator by 10 °C until the baseline noise decreases. |
5 | Nonvolatile additive in mobile phase | Use a volatile mobile phase. |