When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| 3 mm hex key |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7120-68135 | Jet Weaver 35 μL/100 μL |
1 | ![]() | G7120-68380 | Jet Weaver 380 μL (Optional for G7120A) |
1 | ![]() | G7132-68135 | Jet Weaver 35 µL/100 µL biocompatible |
1 | ![]() | G7132-68380 | Jet Weaver 380 µL biocompatible (Optional for G7132A) |
1 | ![]() | G4220-87000 | Capillary ST 0.17 mm x 300 mm Valve to Jet Weaver |
1 | ![]() | G7132-60002 | Biocompatible capillary MP35N 0.17 m x 300 mm Valve to Jet Weaver |
Remove capillary connections from the Jet Weaver.
Remove the hex screws that fix the Jet Weaver to the pump housing.
NOTE
The standard Jet Weavers (G7120-68135 (Jet Weaver 35 μL/100 μL) and G7132-68135 (Jet Weaver 35 µL/100 µL biocompatible) ) have a front and a rear side with different internal volumes (35 / 100 µL) that are optimized for a low delay volume or best mixing performance.
The optional Jet Weaver (G7120-68380 (Jet Weaver 380 μL) , not to be used with Bio pumps, or G7132-68380 (Jet Weaver 380 µL biocompatible) ) is recommended for applications, which are challenging regarding mixing noise (e.g. TFA applications) and has just one side.
Remove the Jet Weaver
Install the new Jet Weaver or flip the Jet Weaver to the other side.
Fix the Jet Weaver with the screws.
Reinstall the capillary connections.
The inlet at the left side of the Jet Weaver is connected to the central port of the pump valve by a capillary (length 300 mm, 0.17 mm i.d.). The outlet at the right side is connected to the autosampler.