When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7120-60029 | SSV Valve Assembly |
2 | ![]() | G4220-60035 | Tubing kit 140 mm, 2/pk SSV to shutoff valve or degassing unit |
Close the shutoff valves.
Remove tubing connections between the SSV and the solvent shutoff valves and between the SSV and the degassing unit inlets.
Push down the SSV panel for removing it.
Remove the connector by pushing up the small clip at the bottom of the connector.
Install a new SSV by inserting the connector and clipping the SSV panel to the module top panel.
Reinstall all tubing connections.