Maintenance parts (G7120A) and Maintenance parts (G7132A) show the main user accessible assemblies of the Agilent 1290 Infinity III High-Speed Pump (G7120A) and Agilent 1290 Infinity III Bio High-Speed Pump (G7132A). These parts can be accessed from the front (simple repairs) and don’t require to remove the pump from the system stack.
The hydraulic path shows the flow connections between these main assemblies.
The recommended interval for preventive maintenance is:
100 L (150 L for Long Life Technology) or 1 year (whichever comes first).
This recommendation is valid for LC instruments on which “typical” applications are running.
A “typical” application can be characterized as follows:
pressure range 100 – 800 bar,
flow rates 0.5 – 3.5 mL/min,
typical solvents used in reversed phase LC.