Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4204-81090 | Quaternary Pump/Flexible Pump Inlet Weaver Assembly |
1 | ![]() | G7131-60090 | Inlet Weaver Assembly Biocompatible |
Prerequisites
|
Open the screw at the bottom of the inlet valve.
Open the fitting at the center of the multi-channel gradient valve (MCGV). Remove the inlet weaver from the MCGV.
Pull the Inlet Weaver out of the Inlet Valve.
Insert the new inlet weaver to the inlet valve. Fix the weaver with the plastic screw.
Fix the fitting of the new inlet weaver to the MCGV.
Reconnect tubings between MCGV and degasser.
base-id: 4133685387