Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 8710-0899 | Screwdriver Pozidrive Shaft |
1 | ![]() | 5023-2502 | Hex driver SW-6.35, slitted |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4204-68035 | Jet Weaver Mixer Kit V35 Quat/Flex |
1 | ![]() | G7131-68035 | Jet Weaver Mixer Kit V35 Flex Biocompatible |
1 | ![]() | G4204-68100 | Jet Weaver Mixer Kit V100 Quat/Flex |
1 | ![]() | G7131-68100 | Jet Weaver Mixer Kit V100 Flex Biocompatible |
1 | ![]() | G4204-68380 | Jet Weaver Mixer Kit V380 Quat/Flex each containing two 5500-1253 |
1 | ![]() | G7131-68380 | Jet Weaver Mixer Kit V380 Flex Biocompatible each containing two 5004-0041 |
2 | ![]() | 5500-1253 | Capillary ST 0.17 mm x 130 mm SX/S Jet Weaver to Multipurpose Valve |
2 | ![]() | 5004-0041 | Capillary MP35N 0.17 mm x 130 mm SI/SX |
Prerequisites
|
Open the screw of the Jet Weaver metal lid.
Remove the metal lid by lifting it up (1) and pulling it out of the front panel (2).
Insert the Jet Weaver to the opening in the front panel (1) and push it down (2).
Fix the Jet Weaver with a screw.
Mount both capillary connections to the Jet Weaver observing the correct orientation (long fitting to Jet Weaver).
Connect the inlet capillary of the Jet Weaver to port 2 of the Multipurpose Valve. Connect the outlet capillary to port 1.
Configure the Jet Weaver as mixer in the user interface.