Error ID: 22469, 22470, 29248
Probable cause | Suggested actions | |
---|---|---|
1 | Seal wash solvent bottle empty | Refill seal wash solvent |
2 | Seal wash tubings not primed | Perform Seal Wash Prime |
3 | Problem in seal wash tubing path | Localize the block or leak Replace inlet filter, if appropriate Replace seal wash tubing, if appropriate Do NOT Recycle Seal Wash. If recycling, the notification will always be present. |
4 | Defect seal wash pump | Replace seal wash pump |
base-id: 3911955339