Error ID: 25471
Sample tray not inserted correctly or missing in container.
Probable cause | Suggested actions | |
---|---|---|
1 | No vial in the position defined in the method. | Install the sample vial in the correct position. Edit the method or sequence accordingly. |
2 | Defective needle assembly. | Exchange the needle assembly. |
3 | Sample container missing or not correctly installed. | Install the sample container correctly on the tray. |
base-id: 3870402571
id: 9007203125143563