The auxiliary pump parameters are pre-set to fit most applications. Adjustments should only be made when required. Auxiliary pump parameters shows the available auxiliary parameters with their default values.
CAUTION
Upon initialization, the pump ignores the Maximum Flow Gradient value.
This can result in a rapid and uncontrolled pressure increase.
Parameter | Limits | Description |
---|---|---|
| 0.1 – 100 mL/min² default: 100 mL/min² | With this parameter flow rate changes can be ramped up and down slowly to avoid pressure shocks to the column. The default value is 100 mL/min² which in fact turns the function off. |
| 20 – 100 µL default: Auto | The volume one pump piston delivers per stroke. In general, a smaller stroke volume results in lower pump ripple. The Auto setting adjusts the strokes dynamically to the lowest possible value. The strokes can be set individually for pump heads A and B. |
| 0 - 150·10-6 /bar or enhanced compressibility calibration default: use enhanced comp. calibration | For best performance, check option Use enhanced compressibility calibration. With this option, the pump will use solvent data libraries provided by Agilent or data generated by using solvent compressibility calibrations. For backward compatibility to 400 bar pumps, the solvent compressibility can still be set manually for each channel when the box is unticked. |
| 0.1 – 100 mL/min² default: 100 mL/min² | With this parameter flow rate changes can be ramped up and down slowly to avoid pressure shocks to the column. The default value is 100 mL/min² which in fact turns the function off. |
| 20 – 100 µL default: Auto | The volume one pump piston delivers per stroke. In general, a smaller stroke volume results in lower pump ripple. The Auto setting adjusts the strokes dynamically to the lowest possible value. The strokes can be set individually for pump heads A and B. |
| 0 - 150·10-6 /bar or enhanced compressibility calibration default: use enhanced comp. calibration | For best performance, check option Use enhanced compressibility calibration. With this option, the pump will use solvent data libraries provided by Agilent or data generated by using solvent compressibility calibrations. For backward compatibility to 400 bar pumps, the solvent compressibility can still be set manually for each channel when the box is unticked. |
base-id: 3836700427
id: 9007203091441419