The pump control parameters are in three sections:
Pump
Seal Wash
Automatic Turn On
Parameter | Limits | Description |
---|---|---|
Pump |
| Enables you to switch the pump On, Off or to a Standby condition. In the Standby condition, the pump motor is still active, and when the pump is switched on again, does not need to be re-initialized. |
Seal Wash |
| The seal wash can be set up to be run once (Single wash) or periodically (Periodic).
The option is available only when the pump has seal wash capability. The seal wash capability is detected by the module itself. If seal wash is installed, it is recommended to use it in order to increase the primary seal lifetime. |
Automatic Turn On |
| Module can be turned on at a specified date/time. This feature can only be used if the module on/off switch is turned on. |
base-id: 3821609995