Error ID: 2201, 2211
Servo Restart Failed: Pump channel A
B: Servo Restart Failed: Pump channel B
The pump motor in the module was unable to move into the correct position for restarting.
When the module is switched on, the first step is to switch on the C phase of the variable reluctance motor. The rotor should move to one of the C positions. The C position is required for the servo to be able to take control of the phase sequencing with the commutator. If the rotor is unable to move, or if the C position cannot be reached, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Mechanical blockage of the module. | |
2 | Disconnected or defective cable. | Please contact your Agilent service representative. |
3 | Blocked (passive or active) inlet valve. | Exchange the (passive or active) inlet valve. |
4 | Defective pump drive assembly. | Please contact your Agilent service representative. |
5 | Defective mainboard. | Please contact your Agilent service representative. |