Error ID: 2043
Valve Fuse 0: Channels A and B
Valve Fuse 1: Channels C and D
The gradient valve in the quaternary pump has drawn excessive current causing the electronic fuse to open.
Probable cause | Suggested actions | |
---|---|---|
1 | Defective gradient valve. | Restart the quaternary pump. If the error message appears again, exchange the gradient valve. |
2 | Defective connection cable (front panel to mainboard). | Please contact your Agilent service representative. |
3 | Defective mainboard. | Please contact your Agilent service representative. |
base-id: 3821124107
id: 9007203075865099