Error ID: 2203, 2213
Index Limit: Pump channel A
B: Index Limit: Pump channel B
The time required by the piston to reach the encoder index position was too short (pump).
During initialization, the first piston is moved to the mechanical stop. After reaching the mechanical stop, the piston reverses direction until the encoder index position is reached. If the index position is reached too fast, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Irregular or sticking drive movement. | Remove the pump head, and examine the seals, pistons, and internal components for signs of wear, contamination or damage. Exchange components as required. |
2 | Defective pump drive assembly. | Please contact your Agilent service representative. |