Error ID: 25236
The tag data of a transport motor cannot be read. The motor ID is given in the event parameter: 0=A, 1=B, 2=Z1, 3=Z2.
Probable cause | Suggested actions | |
---|---|---|
1 | One of the sample handler cables is not properly connected. | Please contact your Agilent service representative. |
2 | One of the sample handler cables is damaged (corroded or chipped off). | Please contact your Agilent service representative. |
3 | Defective mainboard. | Please contact your Agilent service representative. |
base-id: 3803988107
id: 9007203058729099