Error ID: 25105
Probable cause | Suggested actions | |
---|---|---|
1 | Parkstation is loose. | Carefully tighten the parkstation. Avoid overtightening, as this could damage the baseplate of the module. |
2 | Needle assembly is defective. | Replace the needle assembly. |
3 | Autoreferencing needed. | Manually install the needle into the parkstation, clear current autoreferencing values (use Clear data on Lab Advisor), power cycle the module and perform autoreferencing. |