Error ID: 25106
Probable cause | Suggested actions | |
---|---|---|
1 | Autoreferencing values missing or outdated. | Manually install the needle into the parkstation, clear current autoreferencing values (use Clear data on Lab Advisor), power cycle the module and perform autoreferencing. |
base-id: 3803868555