Error ID: 25478
The robot (sample handler) failed to move correctly during injection sequence.
Probable cause | Suggested actions | |
---|---|---|
1 | Missing vessel. | Check if the sample vial is installed in the correct position, or edit the method or sequence accordingly. |
2 | Needle command failed. | Check the status of the needle assembly. Perform an autoreferencing. |