G5667-68711 (Multidraw upgrade kit (Bio-inert)) contains:
|
For bio-inert modules use bio-inert parts only! |
Parts list
p/n | Description | |
---|---|---|
![]() | 5067-4741 | ZDV union (Bio-inert) |
![]() | 0101-1234 | Sample loop 2 mL |
![]() | 0101-1236 | Sample loop 500 µL |
NOTE
The PEEK Sample Loops present in this Multi-draw upgrade kit have lower pressure limit compared to the stainless-steel clad PEEK capillaries installed in other parts of the Bio-Inert Multisampler. The pressure limits depend on the solvent and are listed in the following.
Solvent | Pressure Limit |
---|---|
Acetonitrile |
200 bar |
Water |
240 bar |
Isopropanol |
275 bar |