For bio-inert modules use bio-inert parts only! |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5067-4232 | 2-position/6-port Injection Valve (VICI) 1300 bar (G7167B) |
1 | ![]() | 5067-6698 | 2-position/6-port RC Injection Valve 800 bar (G7167A) |
1 | ![]() | 5067-4260 | 2-position/8-port Injection Valve Dual Needle 1300 bar |
1 | ![]() | 5067-4263 | 2-position/6-port Injection Valve Bio-inert 600 bar for bio inert solution |
1 | ![]() | 5067-6739 | 2-position/6-port injection valve Bio 1300 bar (G7137A) |
Prerequisites
|
NOTE
Please bear in mind that depending on which valve you have installed the images may slightly differ from the actual item.
Disconnect the capillaries.
Turn the spanner nut counter clockwise until the injection valve head detaches from the hydraulic box (Do not use wrenches on the spanner nut).
Remove the spanner nut from the injection valve head.
Take the replacement injection valve head and insert it into the open actuator slot of the hydraulic box. Rotate until the unions at the base of the replacement injection valve head and the valve actuator engage.
OR
If the outside pin does not fit into the outside groove, you have to turn the valve head until you feel that the two pins snap into the grooves. Now you should feel additional resistance from the valve drive while continue turning the valve head until the pin fits into the groove.
NOTE
Check the orientation of the rear side.
Verify the correct position of the Valve TAG.
Continue to rotate until the clocking pin in the injection valve head align with the notch in the housing and press the replacement injection valve head into the actuator.
Replace the spanner nut (1.) and tighten clockwise (2.) (Hand tighten only, do not use wrenches on the spanner nut).
Reconnect the capillaries