Error ID: 34315, 34316, 34317
These, similarly to the Command failed error events, are non-specific error messages and indicate only that something went wrong with initiating the given maintenance procedure. These error messages are usually coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Start of maintenance procedure failed error message.
Error ID: 34315 → Start of needle/seat change procedure failed
Error ID: 34316 → Start of piston change procedure failed
Error ID: 34317 → Start of gripper change procedure failed
Probable cause | Suggested actions | |
---|---|---|
1 | Maintenance procedure failed. | Identify the source error message and follow the instructions provided. |
base-id: 3758470795