When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5065-4445 | Peristaltic pump cartridge |
Prerequisites
|
WARNING
When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can hold health risks.
NOTE
The peristaltic pump cartridge must be replaced as a single entity. The PharMed tubing is not replaceable separately.
Connect the wash port tubing to the left side of the cartridge tubing (use sand paper to get a good grip on the tubing).
Connect the wash solvent line with the plastic fitting to the right side of the cartridge tubing.
base-id: 3716257675