Error ID: 34310
This, similarly to the Command failed error events, is a non-specific error message and indicates only that something went wrong during the cleanup procedure. The cleanup procedure refers to the process, which is executed when an analysis is aborted (for example, due to a missing vial) and the components of the sampler (such as the needle, gripper arm, and the metering device) need to be reset to the initial conditions. As for the Command failed error events, this error message is usually also coupled with other error events, which provide more specific information on the nature of the issue. To identify these source error events, look for entries with the same timestamp as for the Cleanup after abort failed error message.
Probable cause | Suggested actions | |
---|---|---|
1 | Cleanup procedure failed. | Identify the source error message and follow the instructions provided. |