Injector Steps is a diagnostic tool with which the user can walk through the whole injection process while having manual control over the elemental steps. This tool can be especially beneficial during troubleshooting when close observation of each individual step of the injection cycle can be crucial for success.
NOTE
For some commands, there are prerequisite steps that are executed automatically if not already done, see List of injector step commands.
The execution of the Injector Steps tool encompasses the following elementary steps when the Agilent Lab Advisor Software is used:
Select Injector Steps from the list under Service & Diagnostics.
Enter a valid value for the vial location and the injection volume in the Location and Volume fields, respectively.
Select and execute any of the step commands, see List of injector step commands.
Select Back in the bottom right corner to exit the Injector Steps function.
The vial is returned, the metering plunger moves into the home position, and the injection valve switches into the mainpass position.
Command | Action | Prerequisite StepsPrerequisite steps are executed automatically if not already done |
---|---|---|
Valve Bypass | The injection valve switches into the bypass position. |
|
Plunger Home | The metering piston moves into the home position. | Valve Bypass |
Needle Up | The needle arm moves into the upper position. | Valve Bypass |
Vial to Seat | The gripper arm moves the specified vial to the needle seat. |
|
Needle into Sample | The needle arm moves down and the needle penetrates into the vial.
|
|
Draw | The metering device draws the specified volume into the loop capillary.
|
|
Needle Up | The needle arm moves into the upper position, lifting the needle out of the vial. | Valve Bypass |
Vial to Tray | The gripper arm returns the vial into its original position in the vial drawer. | Needle Up |
Needle into Seat | The needle moves down into the needle seat. | Needle Up (only if a vial is present on the needle seat) |
Valve Mainpass | The injection valve switches into the mainpass position. | Needle into Seat |
Reset | The sampler gets reset and will reinitialize |
|
[1]Prerequisite steps are executed automatically if not already done.