The Method dialog box enables the user to change the method parameters relevant for the ICC:
Temperature:
Not Controlled: The temperature of the column oven stays uncontrolled.
At specified temperature (°C): The temperature of the column oven is regulated to stay at the setpoint temperature specified.
NOTE
Valid inputs are 10 to 80 °C in increments of 0.1 °C.
Unchanged: The temperature setting stays unchanged, meaning that the setting currently stored in the system is used by the method.
Valve Position/Column: This section provides information about the location of the column that is being used for the given method.
Enforce column for run: Marking this checkbox restricts the execution of the method to a specific column type, identified by its product number.
Stoptime:
As Pump/Injector: The ICC uses the same stoptime setting as specified for the pump or the sampler.
After specified time (min): The stop time defines a time limit for the analysis, after which all method parameters are reset to initial values.
Posttime:
Off: No post-run period is applied.
For specified time (min): After the analysis, the system stays in the post-run state for the specified time. Using a post-time period can be beneficial when the column needs to be re-equilibrated before starting the next analysis (for example, gradient elutions).
NOTE
For Stoptime and Posttime, valid inputs are 0.01 to 99999 minutes in increments of 0.1 min.
Enable Analysis:
With any temperature: The analysis starts regardless of the actual column oven temperature.
When temperature is within: The analysis starts only when the actual column oven temperature is within the specified range of the setpoint temperature.
Timetable: This feature enables the user to use programmed changes to control the oven temperature during the course of the analysis.