Tools required
Qty. | p/n | Description | |
---|---|---|---|
2 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7130-60030 | Integrated Column Compartment , 3 µL Heater, OR |
1 | ![]() | G7130-60060 | Integrated Column Compartment , 6 µL Heater |
1 | ![]()
|
| Accessory Kit for ICC (provided with the ICC, includes capillaries and column holder clips) |
For more information on recommended capillary types, see Capillary Connections.
Prerequisites
|
WARNING
Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column Compartment are not installed.
NOTE
Plug the ports that are not in use with a blank nut to prevent dust from entering the ICC.
Turn the column onto the Quick Connect Fitting until the first resistance is observed.
Fingertight
Route the capillary between the flap door and the ICC body towards the detector. Alternatively, route the capillary through the incisions at the right side of the ICC.
Turn on the sampler and the pump.
Reconfigure the instrument setup in the CDS.