Load the Default LC Method (def_LC.M).
Select Instrument > Set up Instrument Method...
Under RID set a Stoptime of at least 60 min.
Under Instrument Curves you can select the available diagnostic signals. For baseline issue select the Optical Unit Temperature, Diode Balance, Diode 1 and Diode 2.
Click on Apply and OK.
In the Edit Signal Plot window add the signals shown below.
Set the initial y-range values according to your needs.
The Online Plot Window should look like this:
Save the method as “Test_baseline.M”.
Set up blank runs with 1 – 2 h duration. This will assure that long term effects become visible.
NOTE
Do not run shorter runs than 60 min. Typical effects from air condition systems run at cycles of 15 to 20 min.
Start run when the baseline is stable.