The Standard Optical (G7162A) and Micro Optical (G7162B) detectors described in this manual are designed for highest optical performance, GLP compliance, and easy maintenance. They include the following features:
Advanced temperature controlled detector optics ready to use within two hours of installation
Automatic zero and automatic purge combined with a recycle valve for automatic solvent recycling allow uninterrupted operation
Durable tungsten lamp with a life expectancy of 40000 hours
Automatic light intensity control circuit to ensure the optimum performance of the optics
Integrated diagnostics for efficient troubleshooting
Built-in refractive index calibration
Front access to valves and capillaries for easy maintenance
This detector is equipped with the standard optical unit and is used for all standard applications.
It comprises the following features:
Inlet port to sample cell 62 µL
Inlet port to outlet port 590 µL
Maximum data rate up to 74 Hz
This detector is equipped with the micro optical unit and is used for all low volume applications.
It comprises the following features:
Inlet port to sample cell 2.5 µL
Inlet port to outlet port 265 µL
Maximum data rate up to 148 Hz
LED illumination.