The detector described in this manual is designed for highest optical performance, GLP compliance and easy maintenance. It includes the following features:
120 Hz data acquisition rate for (ultra-) fast LC applications,
RFID tags for all flow cells and UV-lamps provides traceable information about these assemblies
Long-life deuterium with RFID tag and tungsten lamps for highest intensity and lowest detection limit over a wavelength range of 190 – 950 nm
No loss in sensitivity for up to eight wavelengths simultaneous
Programmable slit from 1 – 16 nm for complete optimization of sensitivity, linearity and spectral resolution
Optional flow-cell cartridges with RFID tag (standard 10 mm 13 µL, semi-micro 6 mm 5 µL, micro 3 mm 2 µL, 80 nL, 500 nL, 10 mm, high pressure 10 mm 1.7 µL and prep-cells) are available and can be used depending on the application needs
Easy front access to lamps and flow cell for fast replacement
Built-in holmium oxide filter for fast wavelength accuracy verification
Built-in temperature control for improved baseline stability
Additional diagnostic signals for temperature and lamp voltage monitoring
base-id: 3358299147