NOTE
This information is only valid for UV-lamps without cover (e.g. 2140-0590 and 2140-0813).
Emissions of ultraviolet radiation (200-315 nm) from this product is limited such that radiant exposure incident upon the unprotected skin or eye of operator or service personnel is limited to the following TLVs (Threshold Limit Values) according to the American Conference of Governmental Industrial Hygienists:
Exposure/day | Effective Irradiance |
---|---|
8 hours |
0.1 µW/cm2 |
10 minutes |
5.0 µW/cm2 |
Typically the radiation values are much smaller than these limits:
Position | Effective Irradiance |
---|---|
Lamp installed, 50 cm distance |
Average 0.016 µW/cm2 |
Lamp installed, 50 cm distance |
Maximum 0.14 µW/cm2 |