The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN Interface Card G1369C, and “B”-firmware (B.06.40 or above) or modules with "D"-firmware. All modules should use latest firmware from the same set.
Note the MAC address of the LAN interface (provided with G1369C LAN Interface Card or mainboard). This MAC address is on a label on the card or at the rear of the mainboard, for example, 0030d3177321.
On the Local Controller the MAC address can be found under Details in the LAN section.
Set the configuration switch to DHCP either on the G1369C LAN Interface Card or the mainboard of above mentioned modules.
SW 4 |
SW 5 |
SW 6 |
SW 7 |
SW 8 |
Initialization Mode |
ON |
OFF |
OFF |
OFF |
OFF |
DHCP |
Turn on the module that hosts the LAN interface.
Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition, Lab Advisor) and use MAC address as host name, e.g. 0030d3177321.
The LC system should become visible in the control software (see Note in section General Information (DHCP)).
base-id: 3285103627