When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Screwdriver, Pozidriv #1 PT3 |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G1314-60101 | Deuterium lamp (with RFID tag) |
Prerequisites
|
WARNING
Injury by touching hot lamp
If the detector has been in use, the lamp may be hot.
WARNING
Injury by sharp metal edges
CAUTION
Electronic boards and components are sensitive to electrostatic discharge (ESD).
RFID tag on top. Do not touch the glass bulb with your fingers. It may reduce the light output.
Reset the lamp counter as described in the User Interface documentation (required for non-RFID tag lamps only).
Turn the lamp ON and give the lamp more than 10 minutes to warm-up.
Perform a Wavelength Re-calibration after lamp warm-up.
NOTE
After lamp on, the detector requires a warm-up time of 60 min. No measurements should be performed during this time.
base-id: 3284768651
id: 9007202539509643