Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The parameter of this event equals the measured temperature in 1/100 centigrade. As a result the temperature control is switched off.
Probable cause | Suggested actions | |
---|---|---|
1 | The temperature sensor is defect. | Please contact your Agilent service representative. |
2 | Detector is exposed to illegal ambient conditions. | Verify that the ambient conditions are within the allowed range. |
base-id: 3284348171
id: 9007202539089163