Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed: | Zero-order calibration failed |
Calibration 1 Failed: | 656 nm calibration failed. |
Probable cause | Suggested actions | |
---|---|---|
1 | Lamp is OFF. | Switch on the lamp. |
2 | Incorrect flow cell installation. | Ensure the flow cell is installed correctly. |
3 | Flow cell contamination or air bubbles. | Clean/replace flow cell windows or remove air bubbles. |
4 | Intensity too low. | Replace lamp. |
5 | Current step value too far from maximum. | Repeat the calibration. Please contact your Agilent service representative. |
6 | Misaligned/defective grating assembly. | Please contact your Agilent service representative. |
7 | Defective mainboard. | Please contact your Agilent service representative. |