Error ID: 7318
The holmium oxide test in the detector has failed. During the holmium test, the detector moves the holmium filter into the light path, and compares the measured absorbance maxima of the holmium oxide filter with expected maxima. If the measured maxima are outside the limits, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Misaligned/defective grating assembly. | Ensure the flow cell is inserted correctly, and is free from contamination (cell windows, buffers etc.). Run the filter-motor test to determine if the filter motor assembly is defective. If defective, please contact your Agilent service representative. Run the grating-motor test to determine if the grating assembly is defective. If defective, please contact your Agilent service representative. |