Error ID: 1074
The available power of the heater reached either the upper or lower limit. This event is sent only once per run. The parameter determines which limit has been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause | Suggested actions | |
---|---|---|
1 | Excessive ambient temperature change. | Wait until temperature control equilibrates. |
base-id: 3284195339
id: 9007202538936331